Mts 709 Alignment System User Manual Page 147

  • Download
  • Add to my manuals
  • Print
  • Page
    / 158
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 146
Resetting the FieldPoint Configuration
709 Alignment System Appendix
147
Resetting the FieldPoint Configuration
This section explains how to reset the FieldPoint configuration in case the Align
or Validate tabs cannot connect to the FieldPoint because different FieldPoint
hardware is being used or the configuration was inadvertently changed and
saved.
Refer to the steps in the “FieldPoint Configuration” on page 52 in the Software
Installation section.
1. Locate the Setup Install CD or backup of the correct IAK file (e.g.
CD:\Other\ExtraCopy\Config10.0.0.101.iak).
2. Delete the IAK file in C:\709Align\.
3. Start the National Instruments Measurement & Automation Explorer (NI-
MAX).
4. Expand and delete the Remote Systems in NI-MAX.
5. Restart NI-MAX.
6. Expand the Remote Systems to locate the FieldPoint hardware and setup the
IP address is not already done in NI-MAX (refer to step 2-3).
7. Copy the correct IAK file from the Setup Install CD to C:\709Align\.
8. Open the IAK file in C:\709Align\ with NI-MAX using Tools menu >
FieldPoint > Open (refer to step 4-6).
9. Expand Data Neighborhood and verify data acquisition is functioning (refer
to step 7).
Page view 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 157 158

Comments to this Manuals

No comments