Mts Fracture Template User Manual Page 57

  • Download
  • Add to my manuals
  • Print
  • Page
    / 92
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 56
Caution:
Lower Setpoint/Span Times in the Series 793 Station Manager application can cause
specimen or equipment damage.
Decreasing the Setpoint/Span Times default of 2.0 seconds after the test run completes
can be too fast for other actuator commands, and can cause specimen or equipment
damage.
After the fracture test run has completed, set the Setpoint/Span Times in the Station
Manager application back to their previous values before you use it to command the actuator
for other purposes (for example, for specimen removal or installation).
View Test Results
After the test run completes, you can view the test results.
1. In the Explorer panel, click the name of the test run.
2. Click Results.
3. Click on the different tabs to view different types of results.
Fracture Test Results
The Test Results show the following information about the test run:
Variable Summary shows all the parameters and their last values for the test run.
Crack Size Check Cycle - Data Acquisition shows data that was acquired during the crack size check
activity.
Precrack Command - Data Acquisition shows data that was acquired during the precrack activity.
Data Acquisition shows data that was acquired during the test activity.
The test does not generate any reports. You can generate reports with the results generated during post-test
analysis.
MTS TestSuite | 57
JIC Fracture Toughness Template
Page view 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 91 92