Mts Fracture Template User Manual Page 64

  • Download
  • Add to my manuals
  • Print
  • Page
    / 92
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 63
DescriptionParameter
Specify the percentage of the measured load
range used as the upper limit for linear least
Precrack Upper Least
Squares Fit Percentage
squares fit compliance determination. This
parameter is not used for FCG DCPD tests.
Specify the percent of maximum load level
at which the voltage is measured. The load
Precrack Measure Load
Level Percent
is applied at this level to ensure the crack is
open. All voltage data below this load will not
be used to calculate crack length. This
parameter is used only for FCG DCPD tests.
Specify the number of cycles at which
precrack activity stops.
Precrack Cycle Limit
Specify the maximum stress intensity desired
at the end of precrack activity. The initial
Precrack Final Maximum
K
precrack Maximum K is 1.4 times the Final
Maximum K value.
Select Yes to cause an interlock at the end
of precrack. This can shut down hydraulics,
if desired.
Shutdown Station at End
of Precrack
KIC Parameters
In the KIC Parameters section of the main menu, you specify the KIC Force Ramp Rate, which is the rate
for the ramp segment of the KIC step.
Data Storage Parameters Properties
Data Storage Parameters
DescriptionParameter
Specify the percentage of precrack activity
that must complete before data is saved to
disk.
Precrack Save Percent
Limit
Specify the cycle intervals written to disk
during the precrack activity.
Store Every Nth Precrack
Cycle
Specify the amount of crack length change
that will cause data to be stored to disk.
Crack Size Change Store
64 | MTS TestSuite
KIC Fracture Toughness Template
Page view 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 91 92

Comments to this Manuals

No comments