Mts Fracture Template User Manual Page 81

  • Download
  • Add to my manuals
  • Print
  • Page
    / 92
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 80
DescriptionParameter
Specify the stress intensity (K) value to use
in the load calculation.
Stress Intensity K
Calculate Stress Intensity K
Stress Intensity K Parameters
DescriptionParameter
Specify the crack size value to use in the
stress intensity (K) calculation.
Crack Size
Specify the load value to use in the stress
intensity (K) calculation.
Load
Run the Test
Setpoint/Span Times for Delta-K Control
MTS-supplied fracture test templates calculate setpoint and span values for those portions of a fracture
test that use Delta-K control. For example, the fracture test templates use Delta-K control to pre-crack the
specimen, and the Fatigue Crack Growth template can optionally run in Delta-K control mode.
When a fracture test template uses Delta-K control, it calculates and sends out setpoint and span values
for Delta-K control at a rate of every 1.0 second to the Series 793 Station Manager application. However,
the Station Manager application has default settings for Setpoint/Span Times of 2.0 seconds, which is the
time used to ramp the span and setpoint values. When the Setpoint/Span Times in the Station Manager
application are slower than the setpoint and span update rate in the fracture test template, the controller
may not complete the ramp to a command before the template sends out new span and setpoint corrections
for Delta-K control.
Before running a test with an MTS fracture test template, make sure in the Station Manager application
that the Tools > Channel Options > Command Options > Setpoint/Span Times are set to approximately
0.5 seconds for both Setpoint and Span.
MTS TestSuite | 81
Run the CTOD Fracture Toughness Test
Page view 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 91 92

Comments to this Manuals